4- 76 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
LSSV
The LSSV command specifies and sets the synchronous voltage source for the linear
search measurement. The synchronous source output will be:
Synchronous source output = polarity × LSV source output + offset
where the LSV source output is the value set by the LSV command.
This command setting is cleared by the LSI/LSV command.
Execution
Conditions
The LSV command must be entered before this command.
Syntax LSSV chnum,polarity,offset[,Icomp]
Parameters chnum : Synchronous source channel number. The value must be slot number
where the module has been installed. Integer expression. 1 to 8 (1 and 5
are not available for HPSMU). See
Table 4-1 on page 4-9.
polarity: Polarity of the LSSV output for the LSV output.
0 (negative): LSSV output = −LSV output + offset
1 (positive): LSSV output = LSV output + offset
offset: Offset voltage (in V). Numeric expression.
0 to ±100, or 0 to ±200 for E5280B/E5290A
Both primary and synchronous search sources will use the same output
range. So check the output range set to the LSV command to determine
the synchronous source outputs.
Icomp: Current compliance value (in A). Numeric expression. If you do not
specify Icomp, the previous value is set. Zero amps (0
A) is not a valid
value for the Icomp parameter.
Example
Statements
OUTPUT @E5270;"LSSV 1,0,5,1E-6"
See Also Refer to Table 4-6 on page 4-14 for the source output value, output range, and the
available compliance values.