4- 116 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
TGPC
The TGPC command clears the trigger setting of the specified ports.
Syntax TGPC [port[,port...[,port]...]]
A maximum of 18 ports can be set. If no port is specified, the TGPC command
clears the setting of the all ports; Ext Trig In, Ext Trig Out, and digital I/O ports 1 to
16.
Parameters port : Trigger port number. Integer expression. -1, -2, or 1 to 16.
-1: Ext Trig In terminal.
-2: Ext Trig Out terminal.
1 to 16: Port 1 to 16 of the digital I/O terminal.
Remarks The TGPC command sets the trigger ports as shown below.
Ext Trig In Same as after TGP -1,1,2,0 command execution.
Ext Trig Out Same as after TGP -2,2,2,0 command execution.
Digital I/O Ports No trigger function is available. The ERS? and ERC commands
are available for the port control.
This is not same as the condition set by the *RST command that sets the ports as
shown below.
Ext Trig In Same as after TGP -1,1,2,1 command execution.
Ext Trig Out Same as after TGP -2,2,2,1 command execution.
Digital I/O Ports No trigger function is available. The ERS? and ERC commands
are available for the port control.
Example
Statements
OUTPUT @E5270;"TGPC -1,-2,1,2"
See Also “TGP”

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals