4- 148 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
step : Number of steps for staircase sweep. Numeric expression. 1 to 1001.
The E5260/E5270 can store approximately 16000 measurement data in
ASCII format with header without source data, or approximately 64000
measurement data in binary format.
Icomp : Current compliance (in A). Numeric expression. See Table 4-6 on
page 4-14. If you do not set Icomp, the previous value is used.
Compliance polarity is automatically set to the same polarity as the
output value, regardless of the specified Icomp. If the output value is 0,
the compliance polarity is positive.
If you set Pcomp, the maximum Icomp value for the module is allowed,
regardless of the output range setting.
Pcomp : Power compliance (in W). Numeric expression. Resolution: 0.001 W. If
the Pcomp value is not entered, the power compliance is not set.
0.001 to 2, 0.001 to 4 for E5291A, or 0.001 to 20 for E5280B/E5290A
Example
Statements
OUTPUT @E5270;"WV 1,2,12,1E-6,10,100,0.1,1"
OUTPUT @E5270;"WV 2,1,0,0,20,101"

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals