Keysight E5260/E5270 Programming Guide, Edition 4 4- 149
Command Reference
Command Reference
WZ?
This query command immediately confirms the all channel output, and returns the
status 0 if it is within ±2 V or 1 if it is more than ±2 V.
Syntax WZ? [timeout]
Parameters timeout : Timeout. Numeric expression.
0 to 655.35 sec, with 0.01 sec resolution.
With timeout parameter, this command waits until the all channel
output becomes within ±2 V or until the specified timeout elapses, and
returns 0 or 1.
The WZ? 0 command has the same effect as the WZ? command.
Query Response state<CR/LF^EOI>
0: All channel output is within ±2 V.
1: Any output channel applies more than ±2 V.
Example
Statement
OUTPUT @E5270;"WZ? 5.0"
ENTER
@E5270;A