Keysight E5260/E5270 Programming Guide, Edition 4 5- 7
Error Messages
Error Codes
211 TM1 must be sent to use GET.
Send the TM1 command to use the GPIB GET command (TRIGGER
statement in HP BASIC).
212 Compliance must be set correctly.
Compliance was not set or an incorrect compliance value was set in the
DV, DI, PV, PI, PWV, PWI, TDV, TDI, LSV, LSI, LSSV, LSSI, BSV,
BSI, BSSV, or BSSI command. Set the compliance value correctly.
213 Cannot perform self-test or calibration.
Self-test and calibration cannot be performed in the high voltage state.
Set the output voltage or the voltage compliance to ±42 V or less to
perform the self-test or calibration.
214 Send MM before measurement trigger.
Before sending the measurement trigger, the MM command must be
sent to set the measurement mode.
220 Send WV or WI to set primary sweep source.
Before triggering the staircase sweep measurement, triggering the
staircase sweep with pulsed bias measurement, or sending the WSV,
WSI, or WNX command to set the synchronous sweep source, send the
WV or WI command to set the primary sweep source.
221 Send PWV or PWI to set pulse sweep source.
Before triggering the pulsed sweep measurement, or sending the WSV
or WSI command to set the synchronous sweep source, send the PWV
or PWI command to set the pulse sweep source.
222 Send PV or PI to set pulse source.
Before triggering the staircase sweep with pulsed bias measurement,
send the PV or PI command to set the pulse source.
223 Compliance must be set correctly.
Compliance was not set or an incorrect compliance value was set in the
WV, WI, WSV, WSI, WNX, or BDV command. Set the compliance
value correctly.
224 Sweep and sync output modes must be the same.
The primary sweep channel and the synchronous sweep channel must
be different, and they must be set to the same output mode (voltage or
current).