5- 8 Keysight E5260/E5270 Programming Guide, Edition 4
Error Messages
Error Codes
225 Send WSV, WSI, or WNX to get sync sweep data.
If you enable data output of the synchronous sweep source, do not
forget to set the synchronous sweep source by the WSV, WSI, or WNX
command. For data output, refer to
“FMT” on page 4-59.
226 Set linear sweep for MM4 or MM5.
Only the linear sweep is available for the PWV or PWI command for
the pulsed sweep measurement (MM4) or the WV or WI command for
the staircase sweep with pulsed bias measurement (MM5).
227 Sweep measurement was aborted.
Sweep measurement was aborted by the automatic abort function or the
power compliance.
230 Pulse source must be set.
To perform the pulsed spot measurement (MM3), send the PV or PI
command to set the pulse source.
231 Compliance must be set correctly.
Compliance was not set or an incorrect compliance value was set in the
PV, PI, PWV, or PWI command. Set the correct compliance value
effective for the pulse output.
238 Too large pulse width (max. 2 s).
The maximum value of the pulse width is 2 s. And the available value
depends on the pulse period value. Refer to
“PT” on page 4-90.
239 Pulse width must be 0.5 ms or more.
Set the pulse width to 0.5 ms or more. Refer to “PT” on page 4-90.
253 Program memory is full.
Maximum of 2000 programs or 40000 commands can be stored in the
program memory. Refer to
“ST” on page 4-109.
254 Invalid input for a memory program.
The GPIB GET command (TRIGGER statement in HP BASIC) and an
external trigger input are not allowed in a memory program (between
the ST and END commands).
255 Maximum nesting level is eight.
Nesting (one program calling another) of a memory program must be
eight levels or less.