1- 26 Keysight E5260/E5270 Programming Guide, Edition 4
Programming Basics
Data Output Format
A Status. One character.
Status for Source_data:
Priority of appearance is W<E.
Status for Data, Data_search, or Data_sense: See Table 1-3 on page 1-29.
Priority of appearance is as follows:
For the quasi-pulsed spot measurement: N<T<C<V<X<G or S
For other measurement: N<G<S<T<C<V<X<F
B Channel number of the measurement/source channel. One character.
C Data type. One character.
A Explanation
W Data is for the first or intermediate sweep step.
E Data is for the last sweep step.
B Explanation
A Channel 1.
B Channel 2.
C Channel 3.
D Channel 4.
E Channel 5.
F Channel 6.
G Channel 7.
H Channel 8.
C Explanation
V Voltage measurement data (Data).
I Current measurement data (Data).
T Time data (Time).