Keysight E5260/E5270 Programming Guide, Edition 4 1- 31
Programming Basics
Data Output Format
Staircase Sweep,
Multi Channel
Sweep
Block1 [Block2] . . . . <terminator>
Block1 is the block of data measured at the first sweep step. Block2 is the block of
data measured at the second sweep step.
where Block consists of the following data:
Data1 [Data2] . . . . [Source_data]
DataN (N: integer) is the value measured by a channel. The order of Data is defined
by the MM command.
Source_data is the sweep source output value. It is sent if the data output is enabled
by the FMT command.
Pulsed Sweep,
Staircase Sweep
with Pulsed Bias
Block1 [Block2] . . . . <terminator>
Block1 is the block of data measured at the first sweep step. Block2 is the block of
data measured at the second sweep step.
where Block consists of the following data:
Data [Source_data]
Data is the value measured by the channel you specify by using the MM command.
Source_data is the sweep source output value. It is sent if the data output is enabled
by the FMT command.
Linear Search,
Binary
Search
Data_search [Data_sense]<terminator>
This is the data at the measurement point closest to the search target.
Data_search is the value forced by the search output channel.
Data_sense is the value measured by the search monitor channel. It is sent if data
output is enabled by the BSVM command for the binary search, or the LSVM
command for the linear search.