4 Remote Interface Reference
150 Keysight 34420A User’s Guide
For difference measurements, the specified range applies to the signal connected
to the Channel 1 terminals. Autoranging is selected for voltage measurements on
the Channel 2 terminals.
MEASure:FRESistance?
MEASure:RESistance?
[{<range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX |
DEF}]
This command presets and makes a 2-wire or 4-wire ohms measurement with the
specified range and resolution. The reading is sent to the output buffer.
MEASure:TEMPerature?
{TC | DEF},{B | E | J | K | N | R | S | T | DEF}[,1,{<resolution> |
MIN | MAX | DEF}]
{THER | DEF},{DEF}[,1,{<resolution> | MIN | MAX | DEF}]
{FRTD | DEF},{85 | 91 | DEF}[,1,{<resolution> | MIN | MAX | DEF}]
This command presets and configures the meter for temperature measurements
using the specified probe type and resolution. The reading is sent to the output
buffer. DEF defaults to the last probe type set.
CONFigure[:VOLTage][:DC]
[{<range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX |
DEF}][,(@<channel>)]
This command presets and configures the meter for dc voltage measurements
with the specified range, resolution, and input channel. This command does not
initiate the measurement.
CONFigure[:VOLTage][:DC]:RATio
[{<range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX |
DEF}]
This command presets and configures the meter for dc:dc ratio measurements
with the specified range and resolution. This command does not initiate the
measurement. For ratio measurements, the specified range applies to the signal
connected to the Channel 1 terminals. Autoranging is selected for voltage
measurements on the Channel 2 terminals.