Remote Interface Reference 4
Keysight 34420A User’s Guide 151
CONFigure[:VOLTage][:DC]:DIFFerence
[{<range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX |
DEF}]
This command presets and configures the meter for difference measurements
with the specified range and resolution. This command does not initiate the
measurement. For difference measurements, the specified range applies to the
signal connected to the Channel 1 terminals. Autoranging is selected for voltage
measurements on the Channel 2 terminals.
CONFigure:FRESistance
CONFigure:RESistance
[{<range> | AUTO | MIN | MAX | DEF}][,{<resolution> | MIN | MAX |
DEF}]
This command presets and configures the meter for 2-wire or 4-wire ohms
measurements with the specified range and resolution. This command does not
initiate the measurement.
CONFigure:TEMPerature
{TC | DEF},{B | E | J | K | N | R | S | T | DEF}[,1,{<resolution> |
MIN | MAX | DEF}]
{THER | DEF},{DEF}[,1,{<resolution> | MIN | MAX | DEF}]
{FRTD | DEF},{85 | 91 | DEF}[,1,{<resolution> | MIN | MAX | DEF}]
This command presets and configures the meter for temperature measurements
using the specified probe type and resolution. This command does not initiate the
measurement. DEF defaults to the last probe type set.
CONFigure?
This command queries the meter’s present configuration and return a quoted
string.