5 Error Messages
228 Keysight 34420A User’s Guide
511 RS-232 framing error
512 RS-232 overrun error
513 RS-232 parity error
514 Command allowed only with RS-232
There are three commands which are only allowed with the
RS-232 interface:
SYSTem:LOCal, SYSTem:REMote, and
SYSTem:RWLock.
521 Input buffer overflow
522 Output buffer overflow
531 Insufficient memory
There is not enough memory to store the requested number of
readings in internal memory using the
INITiate command. The
product of the sample count (
SAMPle:COUNt) and the trigger
count (
TRIGger:COUNt) must not exceed 1024 readings.
532 Cannot achieve requested resolution
The meter cannot achieve the requested measurement
resolution. You may have specified an invalid resolution in the
CONFigure or MEASure command.
540 Cannot use overload as math reference
The meter cannot store an overload reading (9.90000000E+37)
as the math reference for null measurements. The math state is
turned off as a result of this condition.
550 Command not allowed in local
The meter received a
READ? command while in the local mode for
RS-232 operation. You should always execute the
SYSTem:REMote command before sending other commands over
the interface.
552 Command not allowed in current function