Chapter 13 245
Assembly Replacement Procedures
Hard Disk Drive
In an effort to eliminate any further confusion for the end user, an additional label
is included in this kit that must be affixed to the rear of the instrument, along with
the Windows Vista license, to explain why the instrument has this Windows Vista
license while running Windows XP.
1. Refer to Figure 13-11. Locate and remove the existing CPU assembly from the
instrument by removing the 6 rear panel screws (1). The CPU assembly can
be removed from the chassis by pulling straight out the back. Use the two
ejectors to pull the CPU assembly out from the chassis.
2. Refer to Figure 13-12. Remove and discard the 15 machine screws (1) from
the CPU assembly. The screws need to be discarded because dried precoated
screws can cause cross threading.
NOTE Do not remove the 4 Hard Drive screws.
Figure 13-12 CPU Top Shield Screws