Chapter 3 85
Instrument Messages
-221 Settings Conflict Errors
-221 Settings conflict; Band is not
available when Mixer Type is
Preselected
The K, E, W, F, D, G, Y, J band functionality is not
available if the selected Mixer Type is Preselected.
-221 Settings conflict; BTS gain is
not available in this Mode
Base Transceiver Station gain correction is not available
in some Modes, or in some measurements (for example,
the SA measurement).
-221 Settings conflict; Calibration
cannot be performed without
valid ENR data
The cal ENR table has no values in it, and hence the cal
cannot be performed. Correct by either populating the
cal ENR table, set ENR mode to Spot, or set the ‘Use
Meas Table Data for Cal’ to ‘On’.
-221 Settings conflict;
Cancellation is not available
while measuring DANL floor.
Phase Noise cancellation does not make sense when
measuring the DANL Floor, so for this reason it has
been disabled
-221 Settings conflict;
Cancellation Ref trace has no
data.
When performing phase noise cancellation, you need to
supply a reference trace that will be used to cancel out
the background noise of the analyzer. The reference
trace must be in Reference (View) mode, and selected
by the Ref Trace parameter under the Cancellation
menu
-221 Settings Conflict;
Cancellation trace has
different X-Scale
Reference trace for the cancellation has a different range
of X-axis against the target trace
-221 Settings conflict; Cannot
optimize while user cal in
progress
Optimize Preselector cannot be performed while a user
cal is in progress. The user cal performs an optimize
preselector prior to taking the noise source on/off level
results for the cal data.
-221 Settings conflict; Carrier
freq not allowed with BMT.
(Bottom/Middle/Top only)
The transmit band spur measurement only allows
bottom (B), middle (M), and top (T) channel
frequencies for each supported frequency band. The
carrier frequency must be set to the bottom, middle or
top frequency of the current frequency band.
-221 Settings Conflict; Code
channel duplication
This error is reported when the given code channel
overlaps other code channel
-221 Settings conflict; Command is
incompatible with band pair
marker
Using remote commands, you have tried to adjust the
start or stop frequency of a span pair marker. You can
only adjust the center and span.
-221 Settings conflict; Continuous
Peak is not available with
Fixed marker
The continuous peak feature cannot be used with a
marker that is fixed. By definition that marker value
cannot change.
-221 Settings conflict; Continuous
Peak is not available with
Signal Track on
The continuous peak feature cannot be used while you
are also using the signal tracking function.