178
Option Esc External Source Control
Using Option ESC with the Keysight MXG Signal Sources.
• By downloading the 70xxxxxx_X.csv file located on the CD ROM disk that is
provided with your mixer. The 70xxxxxx_X.csv file is transferred from the CD to
the USB memory stick (also provided), and then the USB memory stick inserted
into the PXA, and the file transferred to one of the analyzer corrections array
locations.
• The mixer ships with a printed copy of the conversion loss data. Find the printed
copy conversion loss data that has the text "For Use with Keysight X-Series
analyzers only". The conversion loss data will need to be manually entered as
frequency and amplitude pairs into the analyzer corrections file. An example of
calibration data is shown in Figure 14-1 on page 183.
The CD also contains 70xxxxxx_X.pdf files of the conversion loss data that can
be printed, and then manually entered into a Correction array location. You will
need to enter the frequency and amplitude pairs into the analyzer corrections file.
An example of calibration data is shown in Figure 14-1 on page 183.