Chapter 3 Operation
3.13 Factory Default Settings (Initialization)
If you turn the POWER switch on while holding down the RECALL key, all items
listed in
Table 3-15 are set to factory default settings.
Table 3-15 Factory default settings
Item PCR500M PCR1000M PCR2000M
Output Off
Output mode AC
Voltage range 135 V
Voltage AC 0.0 V
DC 0.0 V
Frequency 60.0 Hz
Measured value display RMS
Limit values AC Voltage upper limit 275.0 V
Voltage lower limit 0.0 V
Current limit 5.25 A 10.50 A 21.00 A
Frequency upper limit 500.0 Hz
Frequency lower limit 40.0 Hz
DC Voltage upper limit 388.0 V
Voltage lower limit 0.0 V
Current limit 4.20 A 8.40 A 16.80 A
Current limit operation TRIP
Locking the Panel Operation Released
Configuration OUTPUT on phase OFF
Phase angle 0 deg
Measured value averaging OFF
Peak hold time SHOr (SHORT)
Communication interface type 232 (RS232C)
RS232C baud rate 19.2 (19200 bps)
RS232C XON/XOFF control On
GPIB address 5
Communication error trace function OFF
Memory
A, B, and C
Voltage AC 0.0 V
DC 0.0 V
Frequency 60.0 Hz
Voltage range 135 V
Output mode AC
Limit values
AC
Voltage upper limit 275.0 V
Voltage lower limit 0.0 V
Current limit 5.25 A 10.50 A 21.00 A
Frequency upper limit 500.0 Hz
Frequency lower limit 40.0 Hz
Limit values
DC
Voltage upper limit 385.0 V
Voltage lower limit 0.0 V
Current limit 4.20 A 8.40 A 16.80 A
Current limit operation TRIP