EasyManua.ls Logo

Kikusui PCR 2000M - Page 83

Kikusui PCR 2000M
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PCR-M 83
Appendix
Causes of and remedies for overload status
If the overload protection function is activated, take the remedies below. Then, wait
at least 1 minute before resuming operations.
When the cause of the overload is eliminated, the protection function is
automatically deactivated.
For linear loads
For capacitor-input rectifying loads
For loads that draw an inrush current
Overload Condition Remedy
Gradual output
current increase
If the voltage drops as shown in
Fig. A-4 (a), the current limiting
function is activated.
If the output voltage waveform is
distorted as shown in Fig. A-4 (b-
1) or (b-2), the internal protection
function is activated.
If the current limit value is
set low, change the setting.
If the rated maximum current
is exceeded, reduce the
load.
Sudden output
current increase
If the output voltage waveform is
distorted as shown in Fig. A-4 (b-
1), (b-2), or (c), the internal
protection function is activated.
Overload Condition Remedy
Output current
increase
If the voltage drops as shown in
Fig. A-4 (a), the current limiting
function is activated.
If the current limit value is
set, change the setting.
If the rated maximum current
is exceeded, reduce the
load.
If the voltage waveform is
distorted as shown in Fig. A-4 (c),
the internal protection function is
activated due to the output peak
current.
Reduce the peak current.
Overload Condition Remedy
When voltage is
applied to the load
or when the voltage
suddenly changes
If the voltage waveform is
distorted as shown in Fig. A-4 (c),
the internal protection function is
activated due to the inrush
current.
Reduce the inrush current.

Table of Contents

Related product manuals