9-4 Specifications PLZ-4W
CP mode
*1 It is not applied for the condition of the parrallel operation.
*2 M range applies to the full scale of H range.
Voltmeter
Ammeter
Wattmeter
*1 Displays the product of the voltmeter reading and ammeter reading.
*2 In a mode other the CP mode
*3 In CP mode
Model PLZ164W PLZ334W PLZ1004W PLZ164WA PLZ664WA
Operating range Range H 16.5 W – 165 W 33 W – 330 W 100 W – 1 000 W 16.5 W – 165 W 66 W – 660 W
M 1.65 W – 16.5 W 3.3 W – 33 W 10 W – 100 W 1.65 W – 16.5 W 6.6 W – 66 W
L 0.165 W – 1.65 W 0.33 W – 3.3 W 1 W – 10 W 0.165 W – 1.65 W 0.66 W – 6.6 W
Setting range Range H 0 W – 173.25 W 0 W – 346.5 W 0 W – 1 050 W 0 W – 173.25 W 0 W – 693 W
M 0 W – 17.325 W 0 W – 34.65 W 0 W – 105 W 0 W – 17.325 W 0 W – 69.3 W
L 0 W – 1.7325 W 0 W – 3.465 W 0 W – 10.5 W 0 W – 1.732 5 W 0 W – 6.93 W
Resolution Range H 10 mW 10 mW 100 mW 10 mW 20 mW
M 1 mW 1 mW 10 mW 1 mW 2 mW
L 0.1 mW 0.1 mW 1 mW 0.1 mW 0.2 mW
Accuracy of setting
*1
(0.6 % of set + 1.4 % of f.s
*2
)
Model PLZ164W PLZ334W PLZ1004W PLZ164WA PLZ664WA
Display Range H, M 0.00 V – 150.00 V
L 0.000 V – 15.000 V
Accuracy (0.1 % of rdg + 0.1 % of f.s)
Model PLZ164W PLZ334W PLZ1004W PLZ164WA PLZ664WA
Display Range
H, M
0.000 A
– 33.000 A
0.000 A
– 66.000 A
0.00 A
– 200.00 A
0.000 A
– 33.000 A
0.00 A
– 132.00 A
L
0.00 A
– 330.00 mA
0.00 A
– 660.00 mA
0.0000 A
– 2.0000 A
0.00 A
– 330.00 mA
0.000 A
– 1.3200 A
Accuracy ±(0.2 % of rdg + 0.3 % of f.s)
Parallel operation ±(1.2 % of rdg + 1.1 % of f.s)
Model PLZ164W PLZ334W PLZ1004W PLZ164WA PLZ664WA
Display
*1
Range
H, M
0.00 W
– 165.00 W
0.00 W
– 330.00 W
0.0 W
– 1000.0 W
0.00 W
– 165.00 W
0.00 W
– 660.00 W
L
*2
0.000 W
– 49.500 W
0.000 W
– 99.000 W
0.00 W
– 300.00 W
0.000 W
– 49.500 W
0.000 W
– 198.00 W
L
*3
0.0000 W
– 1.6500 W
0.0000 W
– 3.3000 W
0.000 W
– 10.000 W
0.0000 W
– 1.6500 W
0.0000 W
– 6.6000 W