3. Getting Started
24
Actions
Shaking
Select the Shaking program element if the plate is to be shaken, either before
the measurement or between kinetic cycles.
Enter the respective parameters:
Duration
Enter the duration of the shaking process.
Intensity
Enter the desired shaking mode. Amplitude and frequency
are displayed when choosing the respective shaking mode.
See chapter 4.1.1 for available shaking modes.
Clicking the link Wait a couple of seconds inserts a new program element.
See page
26 for details.
Move Plate
Select the program element Move Plate to move the plate out of or into the
instrument at a certain moment during the workflow.
If the plate is moved out of the reader during a workflow (e.g. to pipette some
liquid into the wells of the microplate), it must be followed by a subsequent
move plate In step, so that the measurement can be finished.