3. Getting Started
26
Miscellaneous
Comment
Use the program element Comment to enter a remark or statement for the
current measurement in the text field.
User Request
The User Request program element informs the operator of the instrument to
execute a definite action during the workflow at a certain time.
If for example the Move Plate program element is used to move the plate out to
perform a certain action, then the entered text should inform the operator to
perform these actions. A dialog box shows the message and the measurement
process stops until OK is clicked.
Wait (Timer)
Use the Wait (Timer) program element to define a certain waiting period before
the next step within a workflow is executed.
In the Wait time field enter the required time.
Enter the respective parameters:
Timer
Enter the Wait time (hh:mm:ss)
Options
Ignore wait at last kinetic cycle: When the program step Wait
(Timer) is the last action within a kinetic run, the wait time will
be ignored in the last cycle.