6.3.1 Interface Commands 55
| www.lakeshore.com
PRBFCOMP? Probe Field Compensation Query
Input PRBFCOMP?[term]
Returned <off/on>[term]
Format n (refer to command for description)
PRBSENS? Probe Sensitivity Query
Input PRBSENS?[term]
Returned <sensitivity>[term]
Format ±nnn.nnnE±nn
Remarks Returns the probe sensitivity in mV/kG
PRBSNUM? Probe Serial Number Query
Input PRBSNUM?[term]
Returned <type>[term]
Format s[8]
Remarks Returns the probe serial number
RANGE Field Range Command
Input RANGE <range>[term]
Format n
<range> Specifies range from lowest to highest: 1 to 4 (field values
are probe dependent)
Example RANGE 4[term]—sets the present range to 4
RANGE? Field Range Query
Input RANGE?[term]
Format n
Returned <range>[term]
Format n (refer to command for description)
RDGFIELD? Field Reading Query
Input RDGFIELD?[term]
Returned <field>[term]
Format ±nnn.nnnE±nn
Remarks Returns the field reading in a format based on the present units—this is valid for
DC or RMS
RDGMODE Measurement Mode Command
Input RDGMODE <mode>,<filter>,<band>[term]
Format n,n,n
<mode> Specifies the measurement mode: 1 = DC, 2 = RMS
<filter> Turns DC filter on or off: 0 = Off, 1 = On
<band> RMS measurement mode: 1 = wide band, 2 = narrow band
Example RDGMODE 2,1,1[term]—the Model 425 is configured for RMS field measurement
in wide band mode