Operation
248083400-1 Rev 7
4-55
When one of the report options buttons is pressed, the instrument uses
the current Room Area, Class Level, Air Flow and Room Status
parameters to determine the minimum number of locations, samples
and volume per channel in order to classify a cleanroom with that
standard.
Use these displayed minimum values to configure the instrument so
that the minimum requirements needed to classify the cleanroom will
be met.
In order to set parameters for a report, follow these steps:
Select the desired report.
Press the SETUP button; the Report Setup screen displays as shown
in Figure 4-68.
Figure 4-68 Report Setup Screen