LightLas 532 Operator's Manual Rev. No 01 Page 38 of 115
3.0 seconds. If the interval is adjusted beyond 3.0 seconds, then the Repeat
Interval display will show “One.
25. Repeat Interval Display
This display will show the selected interval value and it ranges from 0.01 to One
second.
26. Exposure Duration Setting Switches
These two switches control the duration of the exposure time that the
treatment laser is given to the patient each time the footswitch is pressed in the
Treat Mode. The left hand control will decrease the duration and the right hand
control will increase the duration. The minimum duration is 0.01 seconds and the
maximum duration is 3.0 seconds. If the duration is adjusted beyond 3.0
seconds, then the system will enter the continuous pulse mode. In this mode,
the laser will output continuously until the doctor/user releases the footswitch.
When this mode is activated, the Duration display will show “Con”.
27. Exposure Duration Display
The system will store any customized configuration profile at requested time.
This icon is used to store the profile at user.
28. SP Mode & Selection Display Icon
This icon is used to enable and disable SP Mode function and selected sub-
threshold levels will also display. The upper information is the selected duty
cycle and the bottom is the ‘ON’ or exposure time. The denominator is 2000µs
and numerator is appropriate ‘ON’ time. The ratio of the two will be duty cycle.
29. Preset Power Display
This display will show the set power level and the display will change as the
power is adjusted. By default setting the laser power is at set at zero unless the
reconfiguration is made. The setting can be adjustable from 50 mW to 2000 mW
and by rotating the control clockwise will increase the power value and counter-
clockwise will reduce it. The setting will be reset it when power is down.
30. SP Mode Selection Icon
This drop down menu icon will enable the appropriate user preference sub-
threshold duty cycle. There are five preset configurations and they are as follows:
7.5%, 10%, 12.5%, 15%, and 20%. The ‘ON’ time configurations are as follows: 15oµs,
200µs, 250µs, 300µs, 350µs, and 400µs.