LightLas 532 – Operator's Manual Rev. No 01 Page 37 of 115
15. User Configuration Indication
This box will display the user name configuration where appropriate user
preference can be loaded from the folder.
16. Date and Time Indication
This display will show the current time and date. Reconfiguration is required for
system to show the corrected time and date.
17. Total Laser Shot Counter Display
This display will indicate the sum of the total laser shots being fired during the
treatment.
18. Counter Reset Icon
This icon switch will reset the total laser shot counter to zero.
19. Aiming Beam Intensity Level Display
This display will indicate aiming beam intensity level. The level scale starts from
1 to 10 and the higher number indicates the higher intensity. The maximum
power can be applied to patient’s eye is round 1.omW.
20. Aiming Beam Intensity Control Switches
These controls switches can adjust the aiming beam brightness of through two
switches (up / down).
21. Manual Inquiry Icon
This icon will provide the detailed operator manual at any requested time.
22. Loading Customized Configuration Profile Inquiry
This inquiry icon will able to retrieve any customized configuration file.
23. Saving Icon
The system will store any customized configuration profile at requested time.
This icon is used to store the profile at user.
24. Repeat Interval Control Switches
These switches are used to set the repeat interval duration, which is the time
between each pulse of the treatment laser. Similar to the Duration switches, the
left hand control will decrease the interval and the right hand control will
increase the interval. In this mode the repeat pulse rate cannot be set higher
than a 50% duty cycle. This means that the laser fire duration cannot be higher
than the repeat interval duration. It means also that the repeat interval duration
is greater than or equal to the exposure duration when pressing the down
setting switch. The minimum interval is 0.01 second, and the maximum interval is