LightLas 532 – Operator's Manual Rev. No 01 Page 36 of 115
8. Mains Power inlet
This socket is used to connect the Laser Console to the mains power. The input
voltage can be the range of 100 to 230 VAC as the internal PSU is an auto-ranging
module.
9. Footswitch Connector
This connector is fitted for the footswitch module which is the laser trigger firing
mechanism On/Off switch. This footswitch can be in wire or wireless module.
10. Remote Interlock Connector
It is a regulatory requirement that the LightLas 532 have this safety feature
connector available. Two wires can be wired out from the interlock switch to
the treatment door’s switch and once the door opens the circuitry will be
opened causing the system to be disabled. At this time, the warning message
“inL rE-F” will show on the remote control panel display. Once the door shuts
the system will go default back to STANDBY mode. By default setting a dummy
“Interlock” connector is supplied with each console and it must insert in order
for the system to operate correctly.
11. LCD Control Panel Power Switch
This is a power switch for the LCD control panel. By turning on this switch, the
system software will be activated only not the console system itself therefore do
remember to power on the console in order to establish the communication.
12. Power Control Adjuster or Dial
This control is used to adjust the laser output power. By default setting the laser
power is at set at zero unless the reconfiguration is made. The setting can be
adjustable from 50 mW to 2000 mW and by rotating the control clockwise will
increase the power value and counter-clockwise will reduce it. The setting will be
reset it when power is down.
13. Mode Indication (TREAT / STANDBY)
This indication can be toggled the two modes selections of “Treat” or “Standby”.
14. Type of Delivering System Indicator
This display panel will indicate selected type of delivering system when the
appropriate delivery key is inserted or applied. Always ensure the delivery key is
inserted in order for system to function otherwise a warning message “Delivery
Key Not Found” will show in the main LCD panel. There are three types of
delivering systems: Slit Lamp, LIO, and ENDO.