Page #130 background imageLoading...
Page #130 background image
128
MaxiDirect_2c 03/2010
1.1 Methods
Nitrogen, total LR
with Vario Tube Test
0.5 – 25 mg/l N
Insert the adapter for 16 mm Ø vials.
1.
Open two TN Hydroxide LR digestion vials and add
the contents of
one Vario TN Persulfate Rgt. Powder
Pack (Note 2, 3).
2. Add 2 ml deionised water to the prepared vial (this is
the blank, Note 4, 5).
3. Add 2 ml of the water sample to the other prepared vial
(this is the sample).
4. Close the vials with the caps and shake to mix the
contents (at least 30 seconds, Note 6).
5. Heat the vials for 30 minutes in the preheated reactor
at a temperature of 100°C (Note 7).
6. After 30 minutes remove the vials from the reactor.
(CAUTION: The vials are hot!)
Allow the vials to cool to room temperature.
7.
Open the cooled digestion vials and add
the contents of
one
Vario TN Reagent A Powder Pack to each vial (Note 2).
8. Close the vials with the caps and shake to mix the
contents (at least 15 seconds).
9. Press [
] key.
Wait for a reaction period of 3 minutes.
After the reaction period is finished proceed as follows:
10.
Open the digestion vials and add the contents of one
Vario TN Reagent B Powder Pack to each vial (Note 2).
11. Close the vials with the caps and shake to mix the
contents (at least 15 seconds, Note 8).
12. Press [
] key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished proceed as follows:
13. Open two TN Acid LR/HR (Reagent C) vials and
add 2 ml of the digested, treated blank to one vial
(this is the blank).
14. Add 2 ml of the digested, treated water sample to
the other TN Acid LR/HR vial (this is the sample).
15. Close the vials with the caps and swirl the vials gently
several times to mix the contents (10 x, Note 9).
(CAUTION: Vials warm up).
Countdown
3:00
start:
Countdown
2:00
start:
1 2
3 4
Ø 16 mm
8
0
2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lovibond Maxi Direct and is the answer not in the manual?

Lovibond Maxi Direct Specifications

General IconGeneral
BrandLovibond
ModelMaxi Direct
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals