2.3.11 Selecting a new method
Press [ESC] key to return to method selection.
Or enter the required method number directly,
e.g. [Shift] + [1][6][0] for Cyanuric acid.
Confirm with [
] key.
2.3.12 Measure absorbance
Range: –2600 mAbs to +2600 mAbs
Method-No. Title
900 mAbs 430 nm
910 mAbs 530 nm
920 mAbs 560 nm
930 mAbs 580 nm
940 mAbs 610 nm
950 mAbs 660 nm
Select the desired wavelength from the method list or by entering the corresponding method
number directly.
The display shows e.g.:
Always carry out zeroing using a filled (e.g. deionised
water) vial.
The display shows:
Carry out measurement of the sample.
The display shows e.g.:
TIP: To ensure complete reaction times the user countdown may be helpful (chapter 2.2.3,
page 210).
Zero accepted
prepare Test
press TEST
900 mAbs 430 nm
-2600 mAbs - + 2600 mAbs
prepare Zero
press ZERO
500 mAbs
Esc
1 06