1.1 Methods
Suspended Solids
0 – 750 mg/l TSS
Sample preparation:
Blend approx. 500 ml of the water sample in a blender at
high speed for 2 minutes.
1. Fill a clean vial (24 mm Ø) with 10 ml of deionised
water, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the
marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty
the vial completely.
5. Stir the blended water sample. Immediately rinse the
vial with the water sample and fill with 10 ml water
sample.
6. Place the vial in the sample chamber making sure that
the
marks are aligned.
7. Press TEST key.
The result is shown in the display in mg/l TSS (Total
Suspended Solids).
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST