1.1 Methods
Copper, free
with Tablet
0.05 – 5 mg/l Cu
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the
marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one COPPER No. 1 tablet straight from the foil
to the water sample and crush the tablet using a clean
stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that
the
marks are aligned.
8. Press TEST key.
The result is shown in the display in mg/l free Copper.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
5
0
1