215
MaxiDirect_2c 03/2010
Note:
The display shows the number of free data sets.
If there are less than 30 data sets free the display shows:
Clear the memory as soon as possible (see “Deleting stored
results”). If memory capacity is used up it is impossible to
save additional test results.
2.3.9 Printing results (Infra-Red Interface Module) (optional)
If the IRIM (see chapter 2.5) is switched on and the printer is connected, it is possible to print
out the test results (without saving it beforehand).
Press [F3] key.
The entire data set is printed with date, time, Code No.,
method and test result. Printing example:
100 Chlorine T
0.02-6 mg/l Cl
2
Profi-Mode: no
2009-07-01 14:53:09
Test No.: 1
Code-Nr.: 007
4.80 mg/l Cl
2
The test No. is an internal number that is set automatically
if a test result is stored. It appears only on the print out.
2.3.10 Perform additional measurements
To perform additional tests using the same method:
Press [TEST] key
The display shows:
Confirm with [TEST] key
or
Press [ZERO] key to perform a new zero calibration.
The display shows:
Zero accepted
prepare Test
press TEST
prepare Zero
press ZERO
Storage: only 29
free records left
Storage: 900
free records left
F3
Test
Test
Zero