73
MaxiDirect_2c 03/2010
1.1 Methods
Copper, total
with Tablet
0.05 – 5 mg/l Cu
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the
marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one COPPER No. 1 tablet and one COPPER
No. 2 tablet straight from the foil to the water sample
and crush the tablets using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablets are dissolved.
7. Place the vial in the sample chamber making sure that
the
marks are aligned.
8. Press TEST key.
The result is shown in the display in mg/l total Copper.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
5
0
1