EasyManua.ls Logo

Lovibond Maxi Direct - General

Lovibond Maxi Direct
268 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
263
MaxiDirect_2c 03/2010
3.6.2 General
Finding Possible Causes Elimination
Test result deviates from
the expected.
Chemical species not as
required.
Press arrow keys to select the
required chemical species.
No differentiation:
e.g. for the Chlorine test
there is no selection
between differentiated,
free or total.
Profi-Mode is switched on. Switch Profi-Mode off with
Mode 50.
The pre-programmed
countdown is not
displayed.
Countdown is not activated
and/or the Profi-Mode is
activated.
Switch the countdown on
with Mode 13 and/or switch
the Profi-Mode off with
Mode 50.
It seems that a method
is not available.
Method is not activated in
the user method list.
Activate the required method
in the user method list with
Mode 60.

Table of Contents

Related product manuals