1.1 Methods
Hydrazine
with Powder Reagent
0.05 – 0.5 mg/l N
2
H
4
/ 50 – 500 μg/l N
2
H
4
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample (Note 1, 2), close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the
marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add 1 g HYDRAZINE test powder (Note 3) to the
water sample.
6. Close the vial tightly with the cap and swirl several times
to mix the contents.
7. Press [
] key.
Wait for a reaction period of 10 minutes.
After the reaction period is finished proceed as follows:
8. The slight turbidity that occurs when the reagent is added
must be removed by filtration (Note 4).
9. Place the vial in the sample chamber making sure that
the
marks are aligned.
10. Press TEST key.
The result is shown in the display as Hydrazine.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
10:00
start:
0
5
2