EasyManua.ls Logo

Lovibond Maxi Direct - Technical Data

Lovibond Maxi Direct
268 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
259
MaxiDirect_2c 03/2010
3.4 Technical data
Display Graphic Display
Serial Interface IR interface for data transfer
RJ45 connector for internet updates (see chapter 2.5.3)
Light source light-emitting diode – photosensor – pair arrangement
in a transparent measurement chamber
Wavelength ranges:
1 = 530 nm IF = 5 nm
2 = 560 nm IF = 5 nm
3 = 610 nm IF = 6 nm
4 = 430 nm IF = 5 nm
5 = 580 nm IF = 5 nm
6 = 660 nm IF = 5 nm
IF = Interference filter
Wavelength accuracy ± 1 nm
Photometric accuracy* 2% FS (T = 20°C 25°C)
Photometric resolution 0.005 A
Protection conforming to IP 68 (1 h, 0.1 m)
Operation Acid and solvent resistant touch-sensitive keyboard with
integral beeper as acoustic indicator.
Power supply 4 batteries (Type AA/LR 6);
lifetime: approx. 26 hours continuous use or 3500 tests
Auto off 20 minutes after last function,
30 seconds acoustical signal before switch off
Dimensions approx. 210 x 95 x 45 mm (unit)
approx. 395 x 295 x 106 mm (case)
Weight (unit) approx. 450 g
Working condition 5 – 40°C at max. 3090% relative humidity
(without condensation)
Language options English, German, French, Spanish, Italian, Portuguese,
Polish; further languages via Internet Update
Storage capaity ca. 1000 data sets
* measured with standard solutions
Subject to technical modification!
To ensure maximum accuracy of test results, always use the reagent systems
supplied by the instrument manufacturer.