1.1 Methods
Cyanuric acid
with Tablet
2 – 160 mg/l Cys
1. Fill a clean vial (24 mm Ø) with 5 ml of the water
sample and 5 ml deionised water (Note 1), close
tightly with the cap.
2. Place the vial in the sample chamber making sure that
the
marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one CYANURIC ACID tablet straight from the
foil to the prepared water sample and crush the tab-
let using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved (Note 2, 3).
7. Place the vial in the sample chamber making sure that
the
marks are aligned.
8. Press TEST key.
The result is shown in the display in mg/l Cyanuric acid.