1.1 Methods
12. Fill a second clean vial with 10 ml of the water
sample.
13. Add one GLYCINE tablet straight from the foil and
crush the tablet using a clean stirring rod.
14. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
15. Transfer the contents of the second vial into the
prepared vial.
16. Close the vial tightly with the cap and swirl several times
until the tablets are dissolved.
17. Place the vial in the sample chamber making sure that
the
marks are aligned.
18. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l Ozone
mg/l total Chlorine
Notes:
See page 137.
Countdown
2:00
T1 accepted
prepare T2
press TEST
*,** mg/l O
3
*,** mg/l total Cl