1.1 Methods
8. Place the vial in the sample chamber making sure that
the marks are aligned.
9. Press TEST key.
The result is shown in the display in % w/w as availa-
ble chlorine present in the original sample of Sodium
hypochlorite.
Notes:
1. Please pay attention when handling sodium hypochlorite. The material has a very strong
alkalinity and can cause corrosion. Contact with eyes, skin and clothes etc.has to be
avoided. Refer to the detailed information the producer supplied with the product.
2. The tablets must be added in the correct sequence.
3. This method provides a fast and simple test. The test can be performed on site but the
result will not be as precise as a laboratory method.
4. By strictly following the test procedure, an accuracy of +/- 1 weight % can be achieved.
Zero accepted
prepare Test
press TEST