Confirm with [ ] key.
Note:
if the entered number has already been used to save a
polynomial the display shows the query:
• Press [Shift] + [0] or [ESC] keys to go back to method no.
query.
• Press [Shift] + [1] keys to start entry mode.
Enter the required wavelength, e.g.: [2] for 560 nm.
• Press [
] or [ ] key to change between plus and minus
sign
• Enter data of the coefcient A including decimal point,
e.g.: [Shift] + [1][.][3][2]
• Press [F1] key to reset numerical input.
Confirm with [
] key.
• Press [
] or [ ] key to change between plus and minus
sign
•
Enter the exponent of the coefficient A, e.g.: [Shift] + [3]
Confirm with [ ] key.
Successively the instrument queries the data for the other
coefficients (B, C, D, E and F).
Note:
If zero [0] is entered for the value of the coefficient, the
input of the exponent is omitted automatically.
Confirm every input with [
] key.
Enter measurement ranges from – 2600 to +2600 mAbs.
• Press [
] or [ ] key to change between plus and minus
sign.
• Enter the values in Absorbance (mAbs) for the upper limit
(Max) and the lower limit (Min).
Confirm every input with [
] key.
A: 1.32____ E+____
B: +________
measurement range
Min mAbs: +________
Max mAbs: +________
< User polynoms>
y = A+Bx+Cx
2
+Dx
3
+
Ex
4
+Fx
5
A: +________
Overwrite polynom?
YES: 1, NO: 0
wavelength:
1: 530 nm 4: 430 nm
2: 560 nm 5: 580 nm
3: 610 nm 6: 660 nm
1 3 2
.
2
3