EasyManua.ls Logo

Malvern Mastersizer 2000 - Page 61

Malvern Mastersizer 2000
154 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Making measurements - a tutorial Chapter 4
Mastersizer 2000 Page 4-25
Measurement Cycles dialogue
Use the Measurement Cycles dialogue to set up repeat measurements and set up
tank cleaning and dispersant degassing. This diagram shows the features:
ill 7760
Cleaning
It is possible to set up the SOP so that the tank and cell are cleaned together:
Before each aliquot.
After each aliquot.
Before and after each aliquot (select both check boxes).
Selecting the relevant Enable box causes a clean sequence to be performed.
If the sample/dispersant used is not normally removed with a single flush, enter a
number in the Flush cycles box to run the flush sequence that number of times.
The user can set the Clean mode to either Automatic or Manual. Manual ena-
bles the user to clean the system easily using a beaker and a bottle of clean solvent.
A series of prompts guides the user through the cleaning cycle.
Dispersant degassing
Selecting the Enable check box causes a degassing routine to be performed each
time the tank is cleaned. The degassing routine uses ultrasonic action to help
remove dissolved gases from the dispersant.
1
2
MAN0384-1.0 Mastersizer 2000.book Page 25 Tuesday, March 20, 2007 8:56 AM

Table of Contents

Related product manuals