Chapter 6 Understanding the views
Page 6-4 MAN 0384
Report features
A typical report is shown below, followed by a brief description of the key features:.
ill 7755
The residual is an indication of how well the calculated data was fitted to the
measurement data. A good fit is indicated by a residual of under 1%. A residual
of over 1% may indicate use of incorrect refractive index and absorption values
for the sample and dispersant.
The statistics of the distribution are calculated from the results using the
derived diameters D[m,n] - an internationally agreed method of defining the
mean and other moments of particle size. See British standard BS2955:1993 for
more details.
12
3
45
6
7
8
9
10
Dmn,[]
V
i
d
i
m 3
V
i
d
i
n 3
----------------------------
1
mn
-------------
=
MAN0384-1.0 Mastersizer 2000.book Page 4 Tuesday, March 20, 2007 8:56 AM