3-3 Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 2: Basic Setup & Settings
Masterflex
BEFORE USING THE DRIVE
!
CAUTION: Do not block the rear panel of the pump drive. e power switch must always be
easy to access and the power cord must always be easy to disconnect.
!
CAUTION: Turn the drive o before removing or installing pump heads. Fingers or loose
clothing could get caught in the drive mechanism.
!
CAUTION: e power cord set supplied with your pump drive meets the requirements of the
country where you purchased the pump drive. If you use the pump drive in another country,
you must use a power cord set that meets the requirements of that country.
!
CAUTION: If the product is not used in the manner specied in the instructions, the protec-
tion provided by the equipment may be impaired.
CAUTION: e circuit between the mains power supply and the pump must be connected to
earth ground.
CAUTION: Do not handle the pump head before the pump is switched o and disconnected
from the mains power supply.
!
CAUTION: Only use the supplied desktop power supply.
!
CAUTION: Do not run pump head dry.
!
CAUTION: Do not use the gear pump to pump liquids that contain particles. If necessary,
lter the liquid before use.
• Ensure the pump drive is mounted on a at surface.
• Ensure adequate air ow around the pump drive and ensure that the ambient air temperature does not
exceed 104° F (40° C).
• Extension tubing should be clean and routed so that bend radii are at a minimum of four (4) times the
outside tube diameter and are as short as possible.
• Use the appropriate pump head for the required ow rate and viscosity.
• Recalibrate the pump head regularly to maintain the best ow rate accuracy (for further information see
"Pump Head Calibration" on page 3-12).
• For tubing selection and compatibility go to www.masterex.com.
• Remove the pump drive’s power cord from the mains power outlet when cleaning or performing
maintenance on the drive.