3-15 Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 2: Basic Setup & Settings
Masterflex
NOTE:
• e estimated accuracy for the selected calibration settings is shown in the Calibration Strength bar on
the Pump Head Calibration Screen. Calibration accuracy can be improved by using a larger calibration
volume and/or altering the ow rate.
• For the best accuracy, calibration should be performed at the anticipated ow rate.
• Calibration settings are retained and transferred to other mode screens when entering or leaving the Pump
Head Calibration Screen.
• If the drive is stopped during calibration the calibration will be discarded and the previous mode screen
will be displayed. Empty the container before restarting the procedure.
• Calibration time at the maximum allowable ow rate (default max ow rate) is approximately 10 seconds.
Calibration time at the minimum allowable ow rate (approximately 4% of the maximum ow rate) is 4
minutes.
• Minimum and maximum ow rates will change after calibration due to a recalculation of the volume of
uid per pump revolution.
• Steps 8–14 can be repeated as necessary to optimize the accuracy of the pump head calibration.