3-15 Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 2: Basic Setup & Settings
Masterflex
NOTE:
e estimated accuracy for the selected calibration settings is shown in the Calibration Strength bar on
the Pump Head Calibration Screen. Calibration accuracy can be improved by using a larger calibration
volume and/or altering the ow rate.
For the best accuracy, calibration should be performed at the anticipated ow rate.
Calibration settings are retained and transferred to other mode screens when entering or leaving the Pump
Head Calibration Screen.
If the drive is stopped during calibration the calibration will be discarded and the previous mode screen
will be displayed. Empty the container before restarting the procedure.
Calibration time at the maximum allowable ow rate (default max ow rate) is approximately 10 seconds.
Calibration time at the minimum allowable ow rate (approximately 4% of the maximum ow rate) is 4
minutes.
Minimum and maximum ow rates will change after calibration due to a recalculation of the volume of
uid per pump revolution.
Steps 8–14 can be repeated as necessary to optimize the accuracy of the pump head calibration.