12.09.05 rev 1.0
40
46.
Regular Maintenance
The maintenance checks listed in the table below should be carried out as an interim service between the full six
monthly service. It is also valid as quarterly checks in the first twelve months of use. At the end of this period, a full service is
required to the manufacturer’s specification.
Multigen
COMPONENT
VISUAL
OPERATIONAL
NOTES
Power Supply Plugs 9 Check the plug for damage to the pins or strain relief and that all screws are
tight.
Power Supply Cables 9 Check for any damage, abrasion, cuts, or fraying which could compromise the
insulation and the integrity of the power supply cable.
Cable Entry
Connector
9 Check the cable gland nut is tight and the power supply cable is firmly gripped in
the connector.
External Probe 9 9 Check the cord is firmly inserted and secured and the probe tip is not damaged.
ABS Front Panel 9 Check all the retaining screws are in place and not loose. Check the panel has not
been damaged and is held firmly against the seal.
Touch pad Controls 9 9 Check the membrane is not damaged. Ensure all the displays, LED’s and touch
pads function correctly.
Door Latches 9 9 Check the latch for correct operation, loosen the setscrew on the striker plate to
adjust the cone, tighten the screw after any adjustment.
Doors 9 9 Check that they close correctly against the silicon gaskets.
Door Hinges 9 9 Check for excessive wear in the bushings or in the up and down movement of the
door.
Tray rail (Optional) 9 9 Check the hinges are not damaged and that the rail locks in the open position
correctly.
Lateral shelf (Optional) 9 9 Check the support bar screws are tight and that it locates correctly on the end of
the trolley.
Towing Unit Pole End
(Optional)
9 9 Check the spring to ensure it holds the pole at the correct tension so that it does
not drop down and it springs back instantly when pushed down. The spring
must be replaced annually
Towing Unit Female
End (Optional)
9 9 Check the retaining nuts are tight and that the safety pin is still attached.
Castors
9 9 Ensure the brakes function correctly and the wheels are free and not damaged.
Check that the retaining nuts are tight and not entangled with mop strings or
other materials.
Nylon rollers for
transfer module
9 9 Clean and lubricate with food grade silicone spray.
Docking System 9 9 Check that both units align and dock correctly and that the locating pins are not
damaged.
If any faults are found they should be reported immediately and the trolley should be not be used until the fault has been