2.2 Setting up the titration cell for coulometry
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
2.2.4 Titration cell with aspiration equipment (utilization with Dosino)
A Dosino allows the automatic replacement of reagents.
For aspiration, the 6.5617.000 aspiration equipment is used, along with a
complete dosing unit and a 50 mL glass cylinder. For aspirating greasy
samples when only the sample and not the whole reagent is to be aspira-
ted, we recommend using a dosing unit with a 20 mL cylinder. For highly
viscous samples, a dosing unit with a 10 mL cylinder is more suitable.
Mounting the aspiration tip
2
6.1543.070
3
1
6.1446.060
6.2730.030
Insert the aspiration tip into the titration cell as follows:
1
Screw the 6.2730.030 nipple with stopper and O-ring onto the
6.1446.060 stopper.
2
Pull out the stopper.
3
Slide the 6.1543.070 aspiration tip through the stopper.