7 Performing titration with supplementary functions
■■■■■■■■
51
■ Fill out the fields for sample identification and the sample size.
The line number is automatically increased by one after the sam-
ple size has been entered. The next sample size can be entered
directly.
The criterion, after which the line number should be increased by
one, can be modified in the properties dialog of the sample table
under [Editing options].
■ Use the [ ] fixed key to return to the sample table.
■ Switch to the main dialog with the [
] fixed key after the
desired number of sample data has been filled in.
7.4 Carrying out titrations using the sample table
Now carry out the titrations with the samples for which you have entered
sample data.
Each time that you press the [ ] fixed key, the sample data will be loa-
ded from the top line of the sample table and applied for the current titra-
tion. This line will be deleted at the end of the titration. The sample data
of the next line down will be used for the next titration.
A result report will be printed out and a PC/LIMS report will be saved after
each titration if you have defined or activated these options.
NOTICE
The sample table can also be used with automated determinations.