EasyManuals Logo

Metrohm 917 Coulometer User Manual

Metrohm 917 Coulometer
67 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
7.3 Creating a sample table
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
Modifying the settings for the titration
1
Setting an extraction time
NOTICE
The input of an extraction time is recommended, e.g., with sam-
ples that release water slowly or if a Karl Fischer oven (e.g., 885
Compact Oven SC) is used. The extraction time corresponds to the
minimum duration of the titration.
Tap on [Titration parameters].
Under Extraction time, enter the desired time.
Use the [ ] fixed key to return to the KFC command.
2
Modifying the start drift
Tap on [Conditioning].
Under Start drift, enter the desired value.
Define additional stop parameters under [Cond. options].
Conditioning will be canceled if one of the stop parameters set
here is achieved before the start drift is reached.
7.3 Creating a sample table
If you would like to prepare a sample series in advance, you can use the
sample table. Proceed as follows:

Table of Contents

Other manuals for Metrohm 917 Coulometer

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 917 Coulometer and is the answer not in the manual?

Metrohm 917 Coulometer Specifications

General IconGeneral
BrandMetrohm
Model917 Coulometer
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals