7.3 Creating a sample table
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
Modifying the settings for the titration
1
Setting an extraction time
NOTICE
The input of an extraction time is recommended, e.g., with sam-
ples that release water slowly or if a Karl Fischer oven (e.g., 885
Compact Oven SC) is used. The extraction time corresponds to the
minimum duration of the titration.
Tap on [Titration parameters].
Under Extraction time, enter the desired time.
Use the [ ] fixed key to return to the KFC command.
2
Modifying the start drift
Tap on [Conditioning].
Under Start drift, enter the desired value.
Define additional stop parameters under [Cond. options].
Conditioning will be canceled if one of the stop parameters set
here is achieved before the start drift is reached.
7.3 Creating a sample table
If you would like to prepare a sample series in advance, you can use the
sample table. Proceed as follows: