7 Performing titration with supplementary functions
■■■■■■■■
47
NOTICE
The settings for the PC/LIMS report (memory, RS-232, coding)
must be undertaken in the "Edit instrument/PC/LIMS report" dia-
log.
7.2 Modifying the titration parameters
You can optimize a titration run by modifying individual titration parame-
ters according to the needs of your analysis. For the following titration,
the stirring rate, the extraction time and the start drift are being modified.
Setting the stirring rate
1
Opening the stirrer settings
In the main dialog, tap on [Edit parameters].
Select the KFC command and tap on [Edit command].
Tap on [Stirrer].
2
Modifying the stirring rate
Tap on [–] or [+] to modify the stirring rate.
You can use the check box Switch off automatically to set
whether or not the stirrer is switched off at the end of the titra-
tion.
Use the [ ] fixed key to return to the KFC command.
NOTICE
Thorough mixing is important with coulometric titration.
We recommend a Stage 8 stirring rate or Stage 15 when used in
combination with an oven.

Table of Contents

Other manuals for Metrohm 917 Coulometer

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 917 Coulometer and is the answer not in the manual?

Metrohm 917 Coulometer Specifications

General IconGeneral
BrandMetrohm
Model917 Coulometer
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals