7.1 Saving the determination and the PC/LIMS report
■■■■■■■■■■■■■■■■■■■■■■
46
■■■■■■■■
2
Activating Save determination and defining the memory
location
■ Activate the Save determination automatically check box.
■ Select the memory location under Memory. You can choose
between External memory 1 (e.g., a USB flash drive) or Shared
memory (memory location in a computer network).
A USB flash drive (external memory 1) must be plugged in on the
917 Coulometer before a determination is carried out. This mem-
ory location must already be defined in the device manager at the
time of the selection of a shared memory location in a computer
network. Before a determination is carried out, the 917 Coulome-
ter must be connected to the network with an Ethernet cable.
■ Tap on the selection symbol next to the input field Group and
select an existing group or create a new group.
The determination data can be stored on the storage medium in
various groups (= folders on the storage medium).
■ Define the designation for the file name.
Two sample identifications or the method name are available for
selection. If you tap on the input field File name, then you can
enter a designation of your own for the file name.
■ Activate or deactivate write protection.
You can protect the original determination data against overwrit-
ing.
3
Activating the PC/LIMS report and defining the memory loca-
tion
■ Activate the Create PC/LIMS report check box.
Define the memory location for the PC/LIMS report in the device
manager.
■ Return to the main dialog with the [ ] fixed key.
NOTICE
These settings are method-specific. The determination data is
saved to the defined memory location for all determinations which
are carried out with this method. You can define different memory