EasyManua.ls Logo

Michell Instruments CDP301

Michell Instruments CDP301
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CDP301 Dew-Point Tester User’s Manual
Michell Instruments 37
MAINTENANCE
5.3 Mirror Cleaning
Due to contamination in the natural gas, periodic cleaning of the CDP301 mirror may
be required. This must be carried out in a safe area only, in a clean environment.
The cleaning kit supplied contains laboratory cotton buds and a container for cleaning
solvent. Please note that the container is supplied empty due to shipping regulations.
Laboratory-grade Isopropyl alcohol is required for mirror cleaning.
It is recommended that the user also carries the following items:
DSLR camera sensor air blower
Magnifying eyeglass (at least 3x magnication)
A small battery-powered torch
Long-reach 3mm and 4mm Allen keys
Figure 27
Cleaning Accessories
5.3.1 CDP301 Preparation
Ensure that the instrument is powered down and the charging port is not
connected.
Remove all gas pressure from the instrument.
Disconnect all external gas tubing to the ame arrestors

Table of Contents