CDP301 Dew-Point Tester User’s Manual
Michell Instruments 41
MAINTENANCE
5.3.6 Check Mirror Cleanliness
Turn the CDP301 on and allow to boot up.
Turn the CDP301 to Dewpoint Type HCdp measurement and check that
the mirror surface is a uniform colour.
Turn the CDP301 to Dewpoint Type Wdp measurement and check that all
particulates have been removed from the mirror surface.
If the mirror surface is not suitably clean repeat this process.
NOTE: A gas pressure test and appropriate leak check should be carried out in
accordance with local site regulations before the product is used.