EasyManua.ls Logo

MICRODOS Pooltec - Page 32

MICRODOS Pooltec
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation, Programming and Maintenance Microdos Pooltec
Rev. 1.0 EN
32
On device:
1. Press the “
Calibrate” button inside the CL settings menu.
2. Insert the EXPERT PASSWORD (11222) to enter the calibration procedure. The window will appear as the following:
NOTE: it is possible to exit from the calibration procedure simply pushing CANCEL.
3. Press the arrows ▲▼ to insert the value measured by the reference DPD1 test.
Enter the right value then press OK.
4. A window with the value of GAIN obtained from the calibration will appear.
For CL_PB the gain value will be:
For CL_PM the gain value will be:
5. Press
OK to return to the CL menu. The “Last Calibration” box will be updated to the current date.

Table of Contents